With the growing public interest to military and homeland security, a reliable analytical technique for identification and quantitation of explosives is in high demand. Many methods have been proposed including colorimetric assays1 and tandem mass spectrometry2 . EPA has also published a standard method using GC-ECD and GC-MS3 to detect trace explosives. These methods either require a high resolution detector that is not ready for large-scale deployments, or focus on samples in solution. Recently, there are reports4 that established sampling protocols to collect analytes in vapor phase by using sorbent based thermal desorption tubes, followed by a quantitative GC-MS analysis. In this application note, a thermal desorption system from CDS Analytical coupled to mainstream GC-MS was tested on explosive standards spiked in thermal desorption tubes.
Camsco Tenax 1/4” x 3.5” thermal desorption tubes were the sampling device. The thermal desorber was a CDS 7550S 72-position autosampler equipped with Camsco 1/8” x 115 mm focusing trap, which was packed with 80/100 Supelcoport (3% SP-2100) and Tenax sorbent. A Shimadzu QP-2010 GC/MS system with Thermo Scientific TG-SQC capillary column was used as the separation and detection instrument. The optional Peltier cooling module was not equipped for this application. Experimental parameters are listed below:
7550S Thermal Desorber: GC:
Valve oven: 250 °C GC conditions
GC transfer line: 0.01” ID Oven temp.: 40.0 °C
200 °C Injection temp.:
Tube purge flow: 120 mL/min 0 min 160 °C
Pre-heat time: 10 s 2 min 175 °C
Tube desorber: 4 min 270 °C
Rest temp.: 35 °C 10 min 160 °C
Dry purge temp.: 35 °C Injection mode: Split
Dry purge time: 0.5 min Pressure: 30 psi
Desorb temp.: 280 °C Split Ratio: 10.0
Desorb time: 10 min Temp. program:
Trap: 40.0 °C hold 6 min
Rest temp.: 40 °C 120.0 °C ramp to 100.0 °C
Desorb temp.: 250 °C 40.0 °C ramp to 260.0 °C
Desorb time: 4 min hold 10.5 min