Conexiant
Login
  • The Analytical Scientist
  • The Cannabis Scientist
  • The Medicine Maker
  • The Ophthalmologist
  • The Pathologist
  • The Traditional Scientist
The Analytical Scientist
  • Explore

    Explore

    • Latest
    • News & Research
    • Trends & Challenges
    • Keynote Interviews
    • Opinion & Personal Narratives
    • Product Profiles
    • App Notes

    Featured Topics

    • Mass Spectrometry
    • Chromatography
    • Spectroscopy

    Issues

    • Latest Issue
    • Archive
  • Topics

    Techniques & Tools

    • Mass Spectrometry
    • Chromatography
    • Spectroscopy
    • Microscopy
    • Sensors
    • Data and AI

    • View All Topics

    Applications & Fields

    • Clinical
    • Environmental
    • Food, Beverage & Agriculture
    • Pharma and Biopharma
    • Omics
    • Forensics
  • People & Profiles

    People & Profiles

    • Power List
    • Voices in the Community
    • Sitting Down With
    • Authors & Contributors
  • Business & Education

    Business & Education

    • Innovation
    • Business & Entrepreneurship
    • Career Pathways
  • Events
    • Live Events
    • Webinars
  • Multimedia
    • Video
Subscribe
Subscribe

False

The Analytical Scientist / App Notes / 2021 / Analysis of 26 Natural and Synthetic Opioids in Blood and Urine Using Clean Screen® DAU SPE and a Selectra® DA UHPLC Column

Analysis of 26 Natural and Synthetic Opioids in Blood and Urine Using Clean Screen® DAU SPE and a Selectra® DA UHPLC Column

03/08/2021

Share

Featured Image

In recent years, drug abuse has become one of the leading causes of accidental deaths across the country. The opioid crisis was one of the main manifestations of drug-related addictions that caused severe dependency and, in too many cases, fatal overdoses. In this application note, UCT offers a simple yet effective procedure to extract and monitor an extensive panel of opiates in blood and urine using solid-phase extraction (SPE) combined with ultra-high performance liquid chromatography coupled to tandem mass spectrometry (UHPLCMS/MS). Clean Screen® DAU is UCT’s flagship SPE cartridge that can be used to extract a wide range of natural and synthetic opioids with excellent recoveries and overall precision.

In addition, UCT’s Selectra® DA UHPLC column provides excellent retention and peak shape for all the opioids included in the method, including baseline separation of the critical isobaric compounds. Furthermore, all compounds eluted in less than 8 minutes from the Selectra® DA UHPLC column. The simple protocol outlined in this application note can be readily implemented in pain management, clinical diagnostics, and forensic analysis.

Sample Pretreatment:

Urine Specimens*:

  • To 1 mL of urine add 1 mL of pH 6 phosphate buffer (0.1M) and the appropriate volume of internal standard(s)
  • Mix/vortex briefly

٭ A hydrolysis protocol is required if conjugated compounds are added into the drug panel.

Blood Specimens:

  • To 1 mL of blood add 4 mL of pH 6 phosphate buffer (0.1M) and the appropriate volume of internal standard(s)
  • Mix/vortex briefly
  • If necessary (e.g. postmortem blood), centrifuge the sample for 10 minutes at 3000 rpm (discard pellet after loading sample onto SPE cartridge)

>> Download the Full Application Note as a PDF

Newsletters

Receive the latest pathologist news, personalities, education, and career development – weekly to your inbox.

Newsletter Signup Image

Explore More in Pathology

Dive deeper into the world of pathology. Explore the latest articles, case studies, expert insights, and groundbreaking research.

False

Advertisement

Recommended

False

Related Content

Removing User Bias from Structure Verification by NMR
Removing User Bias from Structure Verification by NMR

November 15, 2021

Chemical structure verification by NMR is one of t...

Simultaneous Determination of Eight Nitrosamine Impurities in Metformin Extended-Release Tablets Using the Agilent 6470 Triple Quadrupole LC/MS
Simultaneous Determination of Eight Nitrosamine Impurities in Metformin Extended-Release Tablets Using the Agilent 6470 Triple Quadrupole LC/MS

November 15, 2021

Detection of regulated genotoxic impurities from t...

HIGH RESOLUTION MULTI-REFLECTING TIME-OF-FLIGHT MASS ANALYZER WITH FOLDED FLIGHT PATH®
HIGH RESOLUTION MULTI-REFLECTING TIME-OF-FLIGHT MASS ANALYZER WITH FOLDED FLIGHT PATH®

November 15, 2021

How high does resolving power need to be?

Biomarker discovery
Biomarker discovery

November 16, 2021

Essential guide to analysing VOCs in breath and other biological samples...

False

The Analytical Scientist
Subscribe

About

  • About Us
  • Work at Conexiant Europe
  • Terms and Conditions
  • Privacy Policy
  • Advertise With Us
  • Contact Us

Copyright © 2025 Texere Publishing Limited (trading as Conexiant), with registered number 08113419 whose registered office is at Booths No. 1, Booths Park, Chelford Road, Knutsford, England, WA16 8GS.