Simultaneous Analysis of Saccharides in Non-sugar Beverages Using Single Quadrupole Mass Spectrometer
contributed by Shimadzu |
User Benefits
- The LCMS-2050 can detect saccharides with higher selectivity and sensitivity than differential refractive index detectors or evaporative light scattering detectors.
- The analysis is unaffected by the matrix, allowing simultaneous determination of trace amounts of saccharides.
Introduction
In 2015, the World Health Organization (WHO) issued guidelines that daily sugar intake should be kept below 25g.1) As a standard value stipulated in the Food Labeling Standards of the Consumer Affairs Agency of Japan, a beverage can be labeled as “non-sugar” or “sugar free” if the saccharide content is less than 0.5g per100mL.2)
Sugars are usually detected using a refractive index detector or an evaporative light scattering detector since these compounds have little or no UV absorption. However,if the sugar concentration is low, as in non-sugar beverages, a mass spectrometer with high selectivity and sensitivity is more effective.
This article describes analyses of saccharides in non-sugar beverages separated by hydrophilicinteraction liquid chromatography (HILIC) and detected with a single quadrupole mass spectrometer LCMS-2050.
Analysis of Mixed Standard Solution
The target compounds were nine saccharides that included both monosaccharides to disaccharides.