The separation of standard anions and oxyhalides is performed on the 930 Compact IC Flex instrument after Inline Ultrafiltration. Both anions and oxyhalides can be determined in one single run; the determination complies with EPA 300.1 part A and part B.
Conc [mg/L] n = 4 | RSD [%] n = 4 | Conc. [mg/L] n = 4 | RSD [%] n = 4 | ||
Fluoride | 0.065 | 1.08 | Bromide | 0.004 | 2.12 |
Chlorite | 0.004 | 6.67 | Chlorate | 0.006 | 1.87 |
Bromate | 0.006 | 5.98 | Nitrate | 7.91 | 0.35 |
Chloride | 12.44 | 0.67 | Sulfate | 3.92 | 0.32 |