Fast time-domain near-infrared spectroscopy (TD-NIRS) can noninvasively track oscillations linked to microvascular control in exercising skeletal muscle, according to a proof-of-concept study. The findings highlight the potential of advanced optical spectroscopy to investigate changes in microvascular function associated with aging, exercise training, and disease, offering new opportunities to develop noninvasive methods for physiological assessment and monitoring.
The method combines TD-NIRS measurements acquired at 20 Hz with power spectral density analysis. It measures four hemodynamic parameters: oxyhemoglobin, deoxyhemoglobin, total hemoglobin, and regional muscle oxygen saturation. Analyzing these measurements together could provide more information about muscle microcirculation than monitoring oxygen saturation alone.
Researchers from Politecnico di Milano and the University of L’Aquila tested the approach in eight healthy adults with a mean age of 27 years. Measurements were taken from the tibialis anterior muscle during rest, a 400-second isometric contraction at 40 percent of maximum voluntary contraction, and recovery.
The custom TD-NIRS instrument used picosecond laser sources at 689 and 828 nm, single-photon-sensitive detectors, and time-correlated photon-counting electronics. The optical probe had a source-detector separation of 4 cm, allowing signals to be collected from deeper tissue while reducing interference from skin and subcutaneous fat.
The researchers divided the signals into six frequency bands between 0.005 and 2 Hz. These bands have previously been associated with endothelial, neurogenic, myogenic, respiratory, and cardiac activity, although their precise physiological origins remain under debate.
Spectral patterns differed across rest, contraction, and recovery. Changes were most consistent in the low-frequency bands associated with endothelial, neurogenic, and myogenic regulation. During contraction, regional muscle oxygen saturation fell rapidly, reaching its lowest level after approximately 25 seconds. Total hemoglobin initially decreased before rising, ending around 10 µM above baseline. The researchers interpreted this increase as evidence of greater blood volume and compensatory blood flow.
A superficial blood-volume sensor recorded few significant changes over the same period. This difference suggested that TD-NIRS captured activity in deeper muscle microvasculature that was not clearly visible in measurements dominated by skin perfusion.
Future studies could combine fast TD-NIRS with diffuse correlation spectroscopy, ultrasound, or electromyography and test the method in older adults and people with vascular or muscular disease. Wearable systems may also allow measurements during less restricted movement.