Researchers at University College London have developed reusable surface-enhanced Raman spectroscopy (SERS) substrates by embedding gold or silver nanoparticles within thin diamond films. The approach is intended to protect the plasmonic structures from the physical and chemical damage that can limit the lifetime of conventional SERS substrates.
The study compared silver and gold nanoparticles with median diameters of approximately 10 and 30 nm. The researchers also varied the thickness of the surrounding diamond film to examine how nanoparticle material, particle size, and analyte-to-particle distance affected Raman enhancement.
Metal films were deposited onto single-crystal diamond and converted into nanoparticles through hydrogen-plasma annealing. Diamond was then grown around the particles using microwave plasma chemical vapor deposition. Reactive-ion etching allowed the team to progressively reduce the film thickness and identify configurations that maintained nanoparticle protection while permitting SERS enhancement at the diamond–air interface.
The strongest response came from gold nanoparticles with a median diameter of 27.1 nm encapsulated within a 31.3 nm diamond film. The optimized substrate increased the Raman signal 6,900-fold and detected low concentrations of the test compound. Other gold- and silver-based substrates achieved the same detection limit.
Measurements with methylene blue confirmed that the observed enhancement was not limited to rhodamine 6G.
Gold performed more consistently than silver during substrate fabrication. Diamond grown over silver nanoparticles contained more non-diamond carbon, particularly around the larger particles. The authors attributed this partly to silver nanoparticle movement during high-temperature diamond growth. Gold particles remained more stable and produced higher-quality diamond films across multiple thicknesses.
To assess durability, the researchers subjected the optimized substrates to 30 minutes of sonication, 30 minutes in gold etchant, heating at 400 °C for 30 minutes, and 100 abrasion cycles using 3,000-grit sandpaper. Rhodamine 6G was removed between measurements by solvent sonication. No significant reduction in SERS intensity was observed after the sequential treatments.
The reported detection limit remains higher than those achieved by some highly sensitive single-use SERS platforms. The work instead addresses substrate longevity and repeat use. Further testing with complex samples, target-specific analytes, and larger substrate batches will be needed to establish analytical reproducibility and suitability for routine or field applications.