ADC Analysis with LC-MS compatible mobile phases
contributed by YMC |
Buffers and additives in mobile phase have to be volatile for LC-MS methods, as the solvents are evaporated within the ionisation process. Non-volatile additives would contaminate the mass spectrometer. For small molecules measured at low pH values formic acid is commonly used due to its higher compatibility to MS than TFA. TFA can’t be used in negative mode and remains in the MS. Using formic acid as additive usually produces peak broadening and low intensity for the analysis of proteins.
In this application YMC-Triart Bio C4 was used to analyse a reduced monoclonal antibody (MAb) and a reduced ADC mimic (see Figure 2) with formic acid or TFA as additives. The MAb and ADC mimic were reduced with 10 mM DDT for an hour at 30 °C. The analysis was performed using a YMC-Triart Bio C4 UHPLC column at 80 °C column temperature.